Usage and Characteristics
LDW-80T-Z type drop hammer impact test low temperature chamber is the latest compressor refrigeration equipment developed by our company according to the requirements of GB8363-07 "Temperature Steel Drop Weight Tear Test Method" for cryogenic equipment.
The equipment adopts the compressor refrigeration technology, and uses the principle of heat balance and the stirring circulation mode to achieve automatic uniform cooling and constant temperature of the sample, which can fully meet the various temperature control indexes stipulated by the national standard GB8363-07. The door can be electrically opened and closed, and it can be matched with the manipulator to take the sample.
The equipment is simple and convenient to operate, and has high work efficiency. It is the most ideal sample cooling and heat preservation equipment for metal low temperature impact test. It can also be used for other low temperature testing and testing work.
Main Technical Parameters
2.1, temperature control range: compressor refrigeration: room temperature ~ -80 ° C
2.2, constant temperature accuracy: compressor refrigeration: ± 0.5 ° C
2.3, cooling rate (at room temperature 25 ° C):
Room temperature ~ 0 ° C about 1.3 ° C / min
0°C~ -40°C about 1.0°C/min
-40 ° C ~ -60 ° C about 0.8 ° C / min
-60 ° C ~ -80 ° C about 0.6 ° C / min
2.4, the largest external dimensions: 2280 * 1020 * 1200mm (length * width * height)
2.5, the effective working space: 630 * 380 * 160mm (length * width * height)
2.6, digital timer: 1min ~ 9999min, resolution 1min
2.7. Cooling medium: ethanol or other antifreeze
2.8, stirring motor: 90W
2.9, working power: three-phase five-wire 380V ± 10%, 50Hz, 14kW
2.10, working environment temperature: 5 ° C ~ 25 ° C